Skip to main content
Skip table of contents

Which sequencing platforms are suitable for QuantSeq REV libraries?

Libraries prepared with the QuantSeq 3' mRNA-Seq REV require a Custom Sequencing Primer Version 5 (CSPV5), for Read 1 sequencing. Therefore, REV libraries can be sequenced on platforms that support the use of custom sequencing primers for Read 1.

These platforms include: NextSeq 500 / 550, NextSeq 1000 / 2000, NovaSeq 6000, MiSeq, and AVITI instruments.

QuantSeq REV libraries are not compatible with the iSeq 100 instrument, as this platform does not support the use of custom sequencing primers.

ATTENTION! Always consult the Illumina Custom Sequencing Primer guide or the AVITI Cloudbreak Sequencing User Guide for the relevant instrument.

NOTE: Lexogen does not have direct experience with all of these platforms and reagent versions. These recommendations are based on custom sequencing primer compatibility for these systems. Please inquire at support@lexogen.com for updates, or contact Illumina technical support / Element technical support for further details of custom primer use for your specific sequencing setup.

For loading on Illumina instruments, in general, we recommend loading QuantSeq REV libraries at the higher range of the Illumina-provided specifications.


JavaScript errors detected

Please note, these errors can depend on your browser setup.

If this problem persists, please contact our support.