Is miRVEL Discovery Small RNA-Seq Library Prep Kit compatible with heparinized plasma samples?
Heparin must be removed from RNA samples extracted from heparinized plasma, as heparin is a potent inhibitor of reverse transcription. Treat the isolated RNA with heparinase before subjecting it to miRVEL Discovery Small RNA-Seq Library Prep.